Leica Stereo-Fluorescence SystemsUser Manual

Strany 2

10Leica MZ10 FAssembly 1. Incident- or transmitted-light base 2a. Manual focusing drive, or 2b. Motorized focusing drive 3. Microscope carrier f

Strany 3 - Contents

11Leica MZ10 FFocusing drive ➜ base of stand▶ Connect manually-operated or motor-driven focusing drive, complete with column, to base, as described i

Strany 4 - Safety concept

12AssemblyLamp housingThe Leica 106Z lamp housing is equipped.For information about the supply units for the 106Z lamp housing, see page 14.Model 10

Strany 5

13▶ Using the focusing knob, displace the collector:• Duringthismovement,thecollector must not touch the flexible power cable. If necessary, ben

Strany 6 - Applications

14AssemblySupply unit for lamp housing 106Z with 50W mercury-vapour burner1. On/off switch2. Running-time meter3. Selector for L1/L2 burners4. Fus

Strany 7

15312456Hg50WHg100WAdjusting the mercury vapour lampThis adjustment is important for obtaining a uniform light spot and good-quality fluorescence.

Strany 8

Dummy filter carrierIn the dummy filter carrier there are two openings for the observation beam paths instead of the barrier filters. The third path

Strany 9 - Filter sets for Leica MZ10 F

17UseMZ10 F: Bringing the specimen into focus and observing itObserve the safety directions on page 4 and the recom-mendations on page 5 regarding the

Strany 10 - Leica MZ10 F

18OverviewOutfit with stereo-fluorescence moduleThe components1 Choice of stand and of focusing drive (manual or motor-driven)2 Microscope carrier3

Strany 11

19Filter sets Excitation Dichromatic Barrier filters beam splitters filtersGFP fluorescence 425/60nm 470nm GG475GFP-Plus fluorescence 480/

Strany 13

20AssemblyStereo-fluorescence module 1 Incident- or transmitted-light base 2a Manual focusing drive, or 2b Motorized focusing drive 3 Microscope

Strany 14

21Important: With the illuminator on, insert the filterhol-der with diaphragm into slide-in unit 3 to avoid being suddenly exposed to excessively bri

Strany 15

22AssemblyStereo-fluorescence moduleUV-light excluder ➜ side-faced columnAlways position the UV-light excluder so that the user never looks directly a

Strany 16 - FLUOIII filter system

23UseStereo-fluorescence moduleFocusing and observingObserve the safety regulations on page 4 and note the recommendations on page 5 for using the mer

Strany 17

Leica Microsystems – the brand for outstanding productsLeica Microsystems – an international company with a strong network of customer services.Austra

Strany 18 - Overview

3Contents PageLeica MZ10 F fluorescence stereomicroscopeDescription... 4Safety concept ...

Strany 19

4DescriptionFluorescence techniqueSome substances fluoresce when irradiated with short-wave light. This property is utilized in the fluorescence tech

Strany 20

5Servicing and repairs– Repairs may only be carried out by Leica-trained service technicians or by technical personnel authorized to do so by the per

Strany 21 - Stereo-fluorescence module

6ApplicationsNatural science Application Fluorescence filter setAnatomy •Monitoringofcapillaryflow •GFP1,GFP2Biology •Geneexpressionin

Strany 22

7Industry Application Fluorescence filter setElectronics •SolderpasteonSMDs •GFP1,GFP2 •EpoxyresinonSMDplates •GFP1,GFP2 •Lum

Strany 23 - Focusing and observing

8OverviewLeica MZ10 F: Its components and controls1. Choice of stand and of focusing drive (manual or motor-driven)2. Microscope carrier for Leica M

Strany 24 - Microscopy Systems

9Filter sets Designation Excitation filter Barrier filterGFP fluorescence GFP1 425/60 nm 480 nmGFP Plus fluorescence GFP2 480/40 nm 510 nm GFP